A bootstrap method for gain calibration and resolution determination of a lead-glass calorimeter

  • 1. Department of Physics, University of Connecticut, Storrs, CT 06268 (United States)
  • 2. Department of Physics, Indiana University, Bloomington, IN 47405 (United States)
  • 3. Department of Physics, University of Richmond, Richmond, VA 23173 (United States)
  • 4. Thomas Jefferson National Accelerator Facility, Newport News, VA 23606 (United States)

Description

We describe a method for calibration of a lead-glass calorimeter that does not require a beam of known energy. The technique was used to calibrate the radphi lead-glass calorimeter at Jefferson Lab. The technique described can be applied to any segmented electromagnetic calorimeter capable of detecting all-photon decays of mesons, for example, π0->2γ, η->2γ or ω->π0γ. We also demonstrate how the measured 2γ mass width of the π0 and η mesons can be unfolded to extract the single-shower energy and position resolution functions of the calorimeter