Published January 1992
| Version v1
Journal article
An observed correlation between alpha-particle peak-fitting parameters
- 1. Universidad de Extremadura, Badajoz (Spain). Dept. de Fisica
Description
Using a silicon PIPS detector, a number of alpha-particle spectra were recorded from sources of different nuclide activities. These spectra were fitted using a ''Simplex'' computer programme. The fitting function contains a term which is the convolution of a truncated exponential with a Gaussian function. We have studied the variation of the value of the characteristic energy in the exponential with the value of the variance of the Gaussian. Both values are obtained from the best fits in the least-squares sense. After subtraction of baseline values, we find a significant correlation between the two parameters, which is attributed to the skewed distribution of the energy losses of the alpha-particles prior to detection. (author)
Additional details
Publishing Information
- Journal Title
- Applied Radiation and Isotopes
- Journal Volume
- 43
- Journal Issue
- 1-2
- Series
- Appl. Radiat. Isot.
- Journal Page Range
- 223-227
- ISSN
- 0883-2889
- CODEN
- ARISE
Conference
- Title
- ICRM symposium of low-level-radioactivity measuring techniques and alpha-particle spectrometry.
- Dates
- 4-7 Jun 1991.
- Place
- Monaco (Monaco).
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 23023085
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ALPHA SPECTRA; ALPHA SPECTROMETERS; COMPUTER CALCULATIONS; ENERGY LOSSES; ENERGY RESOLUTION; GAUSS FUNCTION; LEAST SQUARE FIT
- Descriptors DEC
- FUNCTIONS; MAXIMUM-LIKELIHOOD FIT; MEASURING INSTRUMENTS; NUMERICAL SOLUTION; RESOLUTION; SPECTRA; SPECTROMETERS