Published September 1992 | Version v1
Journal article

CAMAC-compatible analog charged particles identifier

  • 1. Inst. of Nuclear Physics, Tashkent (Uzbekistan)

Description

An identification function of the form ΔE(E + 0.472ΔE)0.73 is implemented. In experiments with a beam of 3He particles with energy 25.5 MeV, good mass separation is obtained for five groups of nuclear-reaction products (p, d, t, 3He, and α). Identification dead time is about 3 μsec. 6 refs., 3 figs

Additional details

Publishing Information

Journal Title
Instruments and Experimental Techniques
Journal Volume
35
Journal Issue
2
Journal Page Range
p. 262-265.
ISSN
0020-4412
CODEN
INETAK

Optional Information

Notes
Cover-to-cover Translation of Pribory I Tekhnika Eksperimenta (USSR).