Published August 1994 | Version v1
Journal article

Three-dimensional spiral CT for neurosurgical planning

  • 1. Dept. of Neurosurgery, Technical Univ., Aachen (Germany)
  • 2. Dept. of Neuroradiology, Technical Univ., Aachen (Germany)
  • 3. Dept. of Diagnostic Radiology, Technical Univ., Aachen (Germany)

Description

We carried out 22 examinations to determine the value of three-dimensional (3D) volumetric CT (spiral CT) for planning neurosurgical procedures. All examinations were carried out on a of the first generation spiral CT. A tube model was used to investigate the influence of different parameter settings. Bolus injection of nonionic contrast medium was used when vessels or strongly enhancing tumours were to be delineated. 3D reconstructions were carried out using the integrated 3D software of the scanner. We found a table feed of 3 mm/s with a slice thickness of 2 mm and an increment of 1 mm to be suitable for most purposes. For larger regions of interest a table feed of 5 mm was the maximum which could be used without blurring of the 3D images. Particular advantages of 3D reconstructed spiral scanning were seen in the planning of approaches to the lower clivus, acquired or congenital bony abnormalities and when the relationship between vessels, tumour and bone was important. (orig.)

Additional details

Publishing Information

Journal Title
Neuroradiology
Journal Volume
36
Journal Issue
6
Journal Page Range
p. 435-439.
ISSN
0028-3940
CODEN
NRDYAB