Published February 1, 1998
| Version v1
Report
Temperature analysis for the Hall A cryotarget
Description
In this note the cryogenic data of the Hall A cryotarget are analyzed. The authors will provide a value of the error on the target density. After having checked the consistency of all the temperature and pressure sensors, and corrected them from their offset, the density error is computed. One finds 0.074% for the hydrogen loop and 0.070% density uncertainty for the deuterium loop. A check of the cernox temperature sensors is also provided showing that one cernox in both loop one and loop two has not the same calibration than the two others. All the calibrations for loop three are consistent.
Additional details
Identifiers
Publishing Information
- Imprint Pagination
- 45 Kilobytes
- Report number
- DOE/ER/40150--1253
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 31049415
- Subject category
- S43: PARTICLE ACCELERATORS;
- Resource subtype / Literary indicator
- Non-conventional Literature
- Descriptors DEI
- ACCELERATOR FACILITIES; CEBAF ACCELERATOR; CRYOGENICS; DENSITY; DEUTERIUM; HYDROGEN; PRESSURE MEASUREMENT; TARGETS; TEMPERATURE MEASUREMENT
- Descriptors DEC
- ACCELERATORS; ELEMENTS; HYDROGEN ISOTOPES; ISOTOPES; LIGHT NUCLEI; LINEAR ACCELERATORS; NONMETALS; NUCLEI; ODD-ODD NUCLEI; PHYSICAL PROPERTIES; STABLE ISOTOPES
Optional Information
- Contract/Grant/Project number
- AC05-84ER40150
- Funding organization
- USDOE Office of Energy Research (ER) (United States)
- Secondary number(s)
- JLAB-TN--98-046