Large volume ionization chamber used as laboratory reference for low energy X-ray measurement

  • 1. Institute of Radiation Medicine, Beijing (China)

Description

A large volume spherical ionization chamber of 195 mm diameter and 0.36 mg/cm2 wall thickness made of conducting carbon-fibre epoxy composite material has been developed. The mechanical intensity of the chamber is satisfactory for a good long term volume stability. Owing to its large volume and thin wall, the chamber is sensitive to low energy photon beams and has excellent energy-response characteristics. This ionization chamber is suitable not only for a laboratory reference but also for measurement of low energy photon beam exposure rates at protection-level

Additional details

Publishing Information

Journal Title
Nuclear Science and Techniques
Journal Volume
5
Journal Issue
3
Journal Page Range
p. 166-169.
ISSN
1001-8042
CODEN
NSETEC