Analysis of dosimetric leaf gap variation on dose rate variation for dynamic IMRT

  • 1. Dept. of Radiation Oncology, Chonbuk National University Medical School, Jeonju (Korea, Republic of)

Description

To evaluate the position accuracy of the MLC. This study analyzed the variations of the dosimetric leaf gap(DLG) and MLC transmission factor to reflect the location of the MLC leaves according to the dose rate variation for dynamic IMRT. We used the 6 MV and 10 MV X-ray beams from linear accelerator with a Millennium 120 MLC system. We measured the variation of DLG and MLC transmission factor at depth of 10 cm for the water phantom by varying the dose rate to 200, 300, 400, 500 and 600 MU/min using the CC13 and FC-65G chambers. For 6 MV X-ray beam, a result of measuring based on a dose rate 400 MU/min by varying the dose rate to 200, 300, 400, 500 and 600 MU/min of the difference rate was respectively -2.59, -1.89, 0.00, -0.58, -2.89%. For 10 MV X-ray beam, the difference rate was respectively 2.52, -1.69, 0.00, +1.28, -1.98%. The difference rate of MLC transmission factor was in the range of about 1% of the measured values at the two types of energy and all of the dose rates. This study evaluated the variation of DLG and MLC transmission factor for the dose rate variation for dynamic IMRT. The difference of the MLC transmission factor according to the dose rate variation is negligible, but, the difference of the DLG was found to be large. Therefore, when randomly changing the dose rate dynamic IMRT, it may significantly affect the dose delivered to the tumor. Unless you change the dose rate during dynamic IMRT, it is thought that is to be the more accurate radiation therapy

Additional details

Publishing Information

Journal Title
Journal of the Korean Society for Radiotherapeutic Technology
Journal Volume
28
Journal Issue
1
Series
21 refs, 5 figs, 4 tabs
Journal Page Range
p. 47-55
ISSN
1598-8449

INIS

Country of Publication
Korea, Republic of
Country of Input or Organization
Korea, Republic of
INIS RN
48049878
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
ACCURACY; DOSE RATES; DOSIMETRY; LINEAR ACCELERATORS; NEOPLASMS; RADIOTHERAPY
Descriptors DEC
ACCELERATORS; DISEASES; MEDICINE; NUCLEAR MEDICINE; RADIOLOGY; THERAPY