Tests of an electron monitor for routine quality control measurements of electron energies
Creators
- 1. Department of Radiation Oncology, Health Sciences Center, State University of New York at Stony Brook, Stony Brook, New York (USA)
Description
The depth dose for electrons is sensitive to energy and the AAPM Task Group 24 has recommended that tests be performed at monthly intervals to assure electron beam energy constancy by verifying the depth for the 80% dose to within ±3 mm. Typically, this is accomplished by using a two-depth dose ratio technique. Recently, a new device, the Geske monitor, has been introduced that is designed for verifying energy constancy in a single reading. The monitor consists of nine parallel plate detectors that alternate with 5-mm-thick absorbers made of an aluminum alloy. An evaluation of the clinical usefulness of this monitor for the electron beams available on a Varian Clinac 20 has been undertaken with respect to energy discrimination. Beam energy changes of 3 mm of the 80% dose give rise to measurable output changes ranging from 1.7% for 20-MeV electron beams to 15% for 6-MeV electron beams
Additional details
Publishing Information
- Journal Title
- Medical Physics
- Journal Volume
- 18
- Journal Issue
- 6
- Series
- Med. Phys.
- Journal Page Range
- 1247-1250
- ISSN
- 0094-2405
- CODEN
- MPHYA
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 23008886
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- BEAM MONITORING; DEPTH DOSE DISTRIBUTIONS; ELECTRON BEAMS; ELECTRON DOSIMETRY; ENERGY; MEV RANGE 01-10; MEV RANGE 10-100; QUALITY CONTROL; RADIATION MONITORS; RADIOTHERAPY; TESTING
- Descriptors DEC
- BEAMS; CONTROL; DISTRIBUTION; DOSIMETRY; ENERGY RANGE; LEPTON BEAMS; MEASURING INSTRUMENTS; MEDICINE; MEV RANGE; MONITORING; MONITORS; PARTICLE BEAMS; RADIATION DOSE DISTRIBUTIONS; SPATIAL DISTRIBUTION; SPATIAL DOSE DISTRIBUTIONS; THERAPY