Published January 1996
| Version v1
Report
Open
Calibration of a dosimeter for detection of radon activity in air
Creators
- 1. Dipartimento di Ingegneria Nucleare, Politecnico di Milano, Milan (Italy)
Description
The purpose of this work is to illustrate the calibration of a ionization chamber to be used for the calibration of a cellulose nitrate LRl 15 based dosimeter for radon gas monitoring. During the work the radon was obtain from a source of radium salt of few tens kBq of activity. The calibration factor obtained for the ionization chamber is 3,77xlO15 ± 0.32x1015 Bq/m3 A. For what concern the dosimeter the calibration factor is 4,7xlO-4 ± 1.5x10-4 tr/cm2/Bq·h/m3
Files
37004326.pdf
Files
(234.6 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:b38875acde23b161b33155cab5a274fc
|
234.6 kB | Preview Download |
Additional details
Publishing Information
- Imprint Title
- Symposium on radiation protection in neighbouring countries in Central Europe - 1995. Proceedings
- Imprint Pagination
- 478 p.
- Journal Page Range
- p. 126-130
- Report number
- INIS-XA-C--071
Conference
- Title
- 1995 IRPA regional congress on radiation protection in Central Europe
- Dates
- 4-8 Sep 1995
- Place
- Portoroz (Slovenia)
INIS
- Country of Publication
- International Atomic Energy Agency (IAEA)
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 37004326
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- CALIBRATION; DOSIMETRY; IONIZATION CHAMBERS; RADIATION MONITORING; RADIATION PROTECTION; RADON
- Descriptors DEC
- ELEMENTS; FLUIDS; GASES; MEASURING INSTRUMENTS; MONITORING; NONMETALS; RADIATION DETECTORS; RARE GASES
Optional Information
- Notes
- 1 ref., 4 figs, 3 tabs Imprint:Pages missing in original document