Evaluation of glasses containing cadmium for high dose dosimetry by the thermoluminescence technique

  • 1. Universidade Federal de Uberlandia (INFIS/UFU), MG (Brazil). Instituto de Fisica
  • 2. Instituto de Pesquisas Energeticas e Nucleares (IPEN-CNEN/SP), Sao Paulo, SP (Brazil)
  • 3. Universidade de Sao Paulo (IF/USP), SP (Brazil). Instituto de Fisica

Description

New glass matrices were evaluated for high dose dosimetry by the thermoluminescence technique. Their nominal composition are 20Li2CO3.10Al2O3.15CdO.55B2O3 and 20Li2CO3.10Al2O3.20CdO.50B2O3 (mol%). The glass matrices were irradiated with different doses: 50, 100, 200, 500, 700 and 900 Gy, and the thermoluminescence emission curves were obtained for each of these values. The results show a great potential of using these matrices in high dose dosimetry. (author)

Files

48064536.pdf

Files (232.1 kB)

Name Size Download all
md5:87b4ef158a9cff2889914360bc5ae438
232.1 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
4 p.
Report number
INIS-BR--17781

Conference

Title
3. Brazilian congress on metrology of ionizing radiation
Dates
21-23 Nov 2016
Place
Rio de Janeiro, RJ (Brazil)

INIS

Country of Publication
Brazil
Country of Input or Organization
Brazil
INIS RN
48064536
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
CADMIUM; GLASS SCINTILLATORS; LUMINESCENT DOSEMETERS; QUALITY CONTROL; RADIATION DOSE UNITS; RADIATION PROTECTION; THERMOLUMINESCENCE
Descriptors DEC