Performance evaluation of a mouse-sized camera for dynamic studies in small animals
Creators
- 1. NIMgroup, Biomedical Simulations and Imaging Applications Laboratory, School of Electrical and Computer Engineering, National Technical University of Athens, Iroon Polytecneiou 9, 15780 Zografos (Greece)
- 2. Thomas Jefferson National Accelerator Facility, Newport News, VA 23606 (United States)
- 3. School of Electrical and Computer Engineering, National Technical University of Athens, Iroon Polytecneiou 9, 15780 Zografos (Greece)
- 4. Institute of Radioisotopes-Radiodiagnostic Products, National Center for Scientific Research 'Demokritos', 153 10 Athens (Greece)
Description
A mouse sized camera has been built in terms of collaboration between the presenting institutions. The system is used for the performance of dynamic studies in small animals, in order to evaluate novel radiopharmaceuticals. The active area of the detector is approximately 48x96 mm allowing depiction of the entire mouse in a single view. The system is based on two flat-panel Hamamatsu H8500 position sensitive photomultiplier tubes (PSPMT), a pixellated NaI(Tl) scintillator and a copper-beryllium (CuBe) parallel-hole collimator. In this work, the evaluation results of the system are presented, using phantoms and small animals injected with conventional radiophrmaceuticals. Average resolution was ∼1.6 mm on the collimator surface and increased to ∼4.1 mm in 12 cm distance from the detector. The average energy resolution was measured and found to be ∼15.6% for Tc99m. Results from imaging thin capillaries demonstrated system's high resolution and sensitivity in activity variations was shown. Initial dynamic studies have been carried out in small animals injected with Tc99m-DTPA and Tc99m-MDP. The results show system's ability to perform kinetic imaging in small animals
Additional details
Identifiers
- DOI
- 10.1016/j.nima.2006.10.026;
- PII
- S0168-9002(06)01823-7;
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
- Journal Volume
- 571
- Journal Issue
- 1-2
- Journal Page Range
- p. 48-51
- ISSN
- 0168-9002
- CODEN
- NIMAER
Conference
- Title
- 1. international conference on molecular imaging technology
- Acronym
- EuroMedIm 2006
- Dates
- 9-12 May 2006
- Place
- Marseille (France)
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 38092053
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- BERYLLIUM; CAMERAS; CAPILLARIES; COLLIMATORS; COPPER; DISTANCE; DTPA; ENERGY RESOLUTION; MICE; PERFORMANCE; PHANTOMS; PHOTOMULTIPLIERS; RADIOPHARMACEUTICALS; SENSITIVITY; SODIUM IODIDES; TECHNETIUM 99
- Descriptors DEC
- ALKALI METAL COMPOUNDS; ALKALINE EARTH METALS; AMINO ACIDS; ANIMALS; BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; BLOOD VESSELS; BODY; CARBOXYLIC ACIDS; CARDIOVASCULAR SYSTEM; CHELATING AGENTS; DRUGS; ELEMENTS; HALIDES; HALOGEN COMPOUNDS; HOURS LIVING RADIOISOTOPES; INORGANIC PHOSPHORS; INTERMEDIATE MASS NUCLEI; INTERNAL CONVERSION RADIOISOTOPES; IODIDES; IODINE COMPOUNDS; ISOMERIC TRANSITION ISOTOPES; ISOTOPES; LABELLED COMPOUNDS; MAMMALS; MATERIALS; METALS; MOCKUP; NUCLEI; ODD-EVEN NUCLEI; ORGANIC ACIDS; ORGANIC COMPOUNDS; ORGANS; PHOSPHORS; PHOTOTUBES; RADIOACTIVE MATERIALS; RADIOISOTOPES; RADIOPROTECTIVE SUBSTANCES; RESOLUTION; RESPONSE MODIFYING FACTORS; RODENTS; SODIUM COMPOUNDS; STRUCTURAL MODELS; TECHNETIUM ISOTOPES; TRANSITION ELEMENTS; VERTEBRATES; YEARS LIVING RADIOISOTOPES
Optional Information
- Copyright
- Copyright (c) 2006 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.