Published May 2005 | Version v1
Journal article

Development and calibration of hot electron magnetic spectrometer

  • 1. China Inst. of Atomic Energy, Beijing (China)

Description

The hot electron magnetic spectrometer was developed and its principle, designing idea, structure and index were introduced. The derivation, repeatability, sensitivity and linear absorption coefficient (μ=1.31 mm-1) for LiF TLD were measured by using a 137Cs γ standard source. The spectrometer was calibrated by using a 90Sr β standard source. (author)

Additional details

Publishing Information

Journal Title
Atomic Energy Science and Technology
Journal Volume
39
Journal Issue
3
Journal Page Range
p. 252-256
ISSN
1000-6931

Optional Information

Notes
6 figs., 7 refs.