Published 2004 | Version v1
Miscellaneous

Irradiation dose distribution and safety of 90Sr/90Y prostatic hyperplasia applicator (urethra-type) in the treatment of BPH

  • 1. Department of Nuclear Medicine, The Third Hospital, Jilin University, Changchun (China)

Description

Objective: To study the irradiation dose distribution and safety of 90Sr/90Y prostatic hyperplasia applicator in clinical application. Methods: Characteristics of dosimetry of 90Sr/90Y applicator were determined using nuclear detection techniques in body-phantom. The distribution of the surface dose rate had been measured using lithium fluoride (LiF) thermoluminescent dosimeter during the treatment of the patients. Results: The absorbed doses of β-ray were decreased with the increased distances. The decreased percentages of the absorbed doses at the measured distances of 1 mm, 5 mm and 13 mm to the center of applicator were respectively 57.40%, 99.21% and 99.98% in the body-phantom. No obvious difference was found in the results of surface dose rate of bremsstrahlung. The absorbed doses of doctors and patients were much lower than that of the nation limits. Conclusion: 90Sr/90Y applicator was safe and effective for the treatment of the patients with prostatic hyperplasia. (authors)

Part of:
8th Asia oceania congress of nuclear medicine and biology final program abstracts

Additional details

Publishing Information

Imprint Title
8th Asia oceania congress of nuclear medicine and biology final program abstracts
Imprint Pagination
246 p.
Journal Page Range
p. 126

Conference

Title
8. Asia oceania congress of nuclear medicine and biology
Dates
9-13 Oct 2004
Place
Beijing (China)

Optional Information