SU-G-206-02: Impact of Focal Spot Sizes On CT Image Quality

  • 1. UT MD Anderson Cancer Center, Houston, TX (United States)

Description

Purpose: To quantify a radiology team's assessment of image quality differences between two CT scanner models currently in clinical use, with emphasis on spatial resolution that could be impacted by focal spot size. Methods: Modulation Transfer Functions (MTF) measurements were performed by scanning the impulse source insert module of the Catphan 600 at 120/140 kVp with both large (LFS) and small (SFS) focal spots and reconstructed to 2.5mm and 5.0mm thicknesses on a GE Discovery CT750 HD and a LightSpeed VCT CT scanner. MTFs were calculated by summing the 2D PSF along one-dimension to obtain line-spread-function (LSF), and calculating the Fourier Transform of the zero-padded and background corrected LSF. Spatial resolution performance was evaluated by comparing MTF curves, 50% and 10% MTF cutoff, and total area under the MTF curve (AUC). In addition, images of the Catphan high-contrast module and a Kagaku anthropomorphic body phantom were acquired from the HD scanner for visual comparisons. Results: For each scanner model, SFS was superior to LFS spatial resolution with respect to 50%/10% MTF cutoff and AUC. For the HD, 50%/10% cutoff was 4.29/7.22cm-1 for the LFS and 4.43/7.45cm-1 for the SFS. VCT outperformed HD, with 50%/10% cutoff of 4.40/7.29 cm-1 for LFS and 4.62/7.47cm-1 for SFS. Scanner model performance in order of decreasing AUC performance was VCT SFS (7.43), HD SFS (7.20), VCT LFS (7.09) and HD LFS (6.93). Visual evaluations of Kagaku phantom images confirmed that VCT outperformed HD. Conclusion: VCT outperformed HD and small focal spot is desired for either model over large focal spot in term of spatial resolution – in agreement with radiologist feedback of overall image quality. In-depth evaluations of clinical impact and focal spot selection mechanisms is currently being assessed.

Additional details

Identifiers

Publishing Information

Journal Title
Medical Physics
Journal Volume
43
Journal Issue
6
Journal Page Range
vp.
ISSN
0094-2405
CODEN
MPHYA6

INIS

Country of Publication
United States
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
48097965
Subject category
S60: APPLIED LIFE SCIENCES; S61: RADIATION PROTECTION AND DOSIMETRY;
Descriptors DEI
COMPUTERIZED TOMOGRAPHY; FOURIER TRANSFORMATION; IMAGE PROCESSING; IMAGES; MODULATION; PERFORMANCE; SPATIAL RESOLUTION; TRANSFER FUNCTIONS
Descriptors DEC
DIAGNOSTIC TECHNIQUES; FUNCTIONS; INTEGRAL TRANSFORMATIONS; PROCESSING; RESOLUTION; TOMOGRAPHY; TRANSFORMATIONS

Optional Information

Notes
(c) 2016 American Association of Physicists in Medicine