Published January 1982 | Version v1
Journal article

Evaluation of four assay methods for determination of tobramycin in human serum

  • 1. Univ. of Oklahoma Health Sciences Center, Oklahoma City

Description

Four assay procedures for tobramycin in serum-enzyme immunoassay (I), substrate-labeled fluorescent immunoassay (II), radioimmunoassay (III), and bioassay (IV)-were compared and evaluated by replicate and analytical recovery studies. I and II were about 50% more precise than III and IV. II wassubstantially more nearly accurate than the other methods and also gave the best reproducibility (correlation coefficient 0.992 between-day). The least expensive method was IV. Ease of handling favored I and II. Overall, we find II to be the most acceptable procedure for use in the clinical laboratory

Additional details

Publishing Information

Journal Title
Clin. Chem. (Winston-Salem, N.C.)
Journal Volume
28
Journal Issue
1
Series
Clin. Chem. (Winston-Salem, N.C.).
Journal Page Range
177-180
ISSN