Published 2001 | Version v1
Miscellaneous

LiF(Mg, Cu, P) personal dosimeter intercomparison result and discuss

  • 1. Research Inst. of Chemical Defence, Beijing (China)

Description

The paper introduces the fourth Beijing personal dosimeter compare process and result, the authors used 20 sets of GR-200A thermoluminescent dosimeter was made by solid dosimetry detector and method laboratory, reference radiation adopt 60Co source as National Institute of Metrology, experiment date indicate true measure value is less than 10% of average warp conventional true value, suit for standard request, obtained satisfactory result

Part of:
Proceedings of the 3th national conference on nuclear instrument

Additional details

Publishing Information

Imprint Title
Proceedings of the 3th national conference on nuclear instrument
Imprint Pagination
282 p.
Journal Page Range
p. 261-263

Conference

Title
3. national conference on nuclear instrument
Dates
20-24 Sep 2001
Place
Guilin (China)

Optional Information