Published 2001
| Version v1
Miscellaneous
LiF(Mg, Cu, P) personal dosimeter intercomparison result and discuss
Creators
- 1. Research Inst. of Chemical Defence, Beijing (China)
Description
The paper introduces the fourth Beijing personal dosimeter compare process and result, the authors used 20 sets of GR-200A thermoluminescent dosimeter was made by solid dosimetry detector and method laboratory, reference radiation adopt 60Co source as National Institute of Metrology, experiment date indicate true measure value is less than 10% of average warp conventional true value, suit for standard request, obtained satisfactory result
Additional details
Publishing Information
- Imprint Title
- Proceedings of the 3th national conference on nuclear instrument
- Imprint Pagination
- 282 p.
- Journal Page Range
- p. 261-263
Conference
- Title
- 3. national conference on nuclear instrument
- Dates
- 20-24 Sep 2001
- Place
- Guilin (China)
INIS
- Country of Publication
- China
- Country of Input or Organization
- China
- INIS RN
- 36021525
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference, Non-conventional Literature
- Descriptors DEI
- CALIBRATION; DOSE EQUIVALENTS; INTERLABORATORY COMPARISONS; LITHIUM FLUORIDES; PERSONNEL DOSIMETRY; PERSONNEL MONITORING; THERMOLUMINESCENT DOSEMETERS
- Descriptors DEC
- ALKALI METAL COMPOUNDS; DOSEMETERS; DOSIMETRY; FLUORIDES; FLUORINE COMPOUNDS; HALIDES; HALOGEN COMPOUNDS; LITHIUM COMPOUNDS; LITHIUM HALIDES; LUMINESCENT DOSEMETERS; MEASURING INSTRUMENTS; MONITORING; RADIATION MONITORING