A possibility of parallel and anti-parallel diffraction measurements on neutron diffractometer employing bent perfect crystal monochromator at the monochromatic focusing condition
- 1. HANARO Center, Korea Atomic Energy Research Institute, Daejeon (Korea, Republic of)
- 2. Nuclear Physics Institute, 25068 Rez (Czech Republic)
Description
In a conventional diffractometer having single monochromator, only one position, parallel position, is used for the diffraction experiment (i.e. detection) because the resolution property of the other one, anti-parallel position, is very poor. However, a bent perfect crystal (BPC) monochromator at monochromatic focusing condition can provide a quite flat and equal resolution property at both parallel and anti-parallel positions and thus one can have a chance to use both sides for the diffraction experiment. From the data of the FWHM and the Δd/d measured on three diffraction geometries (symmetric, asymmetric compression and asymmetric expansion), we can conclude that the simultaneous diffraction measurement in both parallel and anti-parallel positions can be achieved. (author)
Additional details
Publishing Information
- Journal Title
- Pramana
- Journal Volume
- 63
- Journal Issue
- 1
- Journal Page Range
- p. 175-181
- CODEN
- PRAMCI
Conference
- Title
- conference on neutron scattering - Pt. 1
- Acronym
- CNS-04
- Dates
- 2-4 Jan 2004
- Place
- Mumbai (India)
INIS
- Country of Publication
- India
- Country of Input or Organization
- India
- INIS RN
- 36032512
- Subject category
- S75: CONDENSED MATTER PHYSICS, SUPERCONDUCTIVITY AND SUPERFLUIDITY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- BRAGG REFLECTION; ENERGY RESOLUTION; FOCUSING; MONOCHROMATORS; NEUTRON DIFFRACTION; SILICON; SPATIAL RESOLUTION
- Descriptors DEC
- COHERENT SCATTERING; DIFFRACTION; ELEMENTS; REFLECTION; RESOLUTION; SCATTERING; SEMIMETALS
Optional Information
- Notes
- 11 refs., 4 figs.