Validation of efficiency calibration of HPGe well-type detector using a 85Sr standard solution
- 1. Institute of Physics, Vilnius (Lithuania)
Description
The application of near 4 pi counting geometry for determining low-level gamma-emitters improves detection limits considerably. At the same time, true coincidence effect is pronounced in a well-type counter, and requires correction for loss of coincidence. Therefore, only single-photon emitting radionuclides can be used for efficiency calibration. An efficiency calibration was performed for a particular HPGe well-type detector using reference standard solutions with 57Co, 137Cs, 54Mn, 65Zn and 40K. This set of radionuclides allowed to construct a calibration curve that provided overall uncertainty of 6% in activity measurements in the energy region of 122-1461 keV. However, the first half of this region (122-662 keV) lacked suitable calibration points; it was therefore expected that any additional point could enhance calibration. 85Sr reference standard solution was applied for this task where available. The counting efficiency was investigated taking into account the filling height of the measuring container. Detector response at 514 keV was included in a new efficiency curve construction, which made it possible to improve the precision of previous calibration.(author)
Files
36052801.pdf
Files
(272.4 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:52d4fef315c3099ce8334422906e9c80
|
272.4 kB | Preview Download |
Additional details
Additional titles
- Original title (English)
- Zbornik radova Sestog simpozija Hrvatskog drustva za zastitu od zracenja
Publishing Information
- Imprint Place
- Zagreb (Croatia)
- ISBN
- 953-96133-5-3
- Imprint Title
- Proceedings of the Sixth Symposium of the Croatian Radiation Protection Association
- Imprint Pagination
- 498 p.
- Journal Page Range
- p. 128-133
- Report number
- INIS-HR--05003
Conference
- Title
- 6. Symposium of the Croatian Radiation Protection Association
- Original Conference Title
- Sesti simpozij Hrvatskog drustva za zastitu od zracenja
- Dates
- 18-20 Apr 2005
- Place
- Zagreb (Croatia)
INIS
- Country of Publication
- Croatia
- Country of Input or Organization
- Croatia
- INIS RN
- 36052801
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- CALIBRATION; CESIUM 137; COBALT 57; EFFICIENCY; GAMMA SOURCES; MANGANESE 54; MEASURING INSTRUMENTS; POTASSIUM 40; ZINC 65
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; BETA-PLUS DECAY RADIOISOTOPES; CESIUM ISOTOPES; COBALT ISOTOPES; DAYS LIVING RADIOISOTOPES; ELECTRON CAPTURE RADIOISOTOPES; EVEN-ODD NUCLEI; INTERMEDIATE MASS NUCLEI; ISOMERIC TRANSITION ISOTOPES; ISOTOPES; LIGHT NUCLEI; MANGANESE ISOTOPES; NANOSECONDS LIVING RADIOISOTOPES; NUCLEI; ODD-EVEN NUCLEI; ODD-ODD NUCLEI; POTASSIUM ISOTOPES; RADIATION SOURCES; RADIOISOTOPES; YEARS LIVING RADIOISOTOPES; ZINC ISOTOPES
Optional Information
- Notes
- 9 refs., 3 tabs., 2 figs. Imprint:Zbornik radova Sestog simpozija Hrvatskog drustva za zastitu od zracenja