There is a newer version of the record available.

Published February 2020 | Version v1
Journal article

Results of a phase I, non-randomized study evaluating a Magnetic Occult Lesion Localization Instrument (MOLLI) for excision of non-palpable breast lesions

  • 1. Sunnybrook Odette Cancer Centre. Department of Surgery (Canada)
  • 2. Sunnybrook Odette Cancer Centre. Department of Medical Physics (Canada)
  • 3. Sunnybrook Research Institute. Clinical Research Program (Canada)

Description

Purpose

: Magnetic Occult Lesion Localization Instrument (MOLLI) is a wireless, non-radioactive alternative for non-palpable breast lesion localization. The primary objective of this first-in-human study was to evaluate the clinical feasibility of using MOLLI for intraoperative localization of non-palpable breast lesions.

Methods

: Twenty women with non-palpable breast lesions at a single institution received a lumpectomy using the MOLLI guidance system. Patients were co-localized with magnetic and radioactive markers up to 7 days before excision by a dedicated breast radiologist under sonographic guidance. Both markers were localized intraoperatively using dedicated hand-held probes. The primary outcome was successful excision of the magnetic marker, confirmed radiographically and pathologically. Demographic data, margin positivity, and re-excision rates were collected. Surgical oncologists, radiologists, and pathology staff were surveyed for user satisfaction.

Results

: Post-radiological analysis: Post-implant mammograms verified that 17/20 markers were placed directly in the lesion center. Radiologists reported that all marker implantations procedures were "easy" or "very easy" following a single training session. Post-surgical analysis: All MOLLI markers were successfully removed with the specimen during surgical excision. In all cases, surgeons ranked the MOLLI guidance system as "very easy" for lesion localization. Pathologic analysis: All patients had negative margins. All anatomic pathology staff ranked the MOLLI system as "very easy" to localize markers.

Conclusions

: The MOLLI guidance system is a reliable and accurate method for intraoperative localization of non-palpable breast lesions. Further evaluation of the MOLLI system in studies against current standards of care is required to demonstrate system cost-effectiveness and improved patient-reported outcomes.

Additional details

Identifiers

Publishing Information

Journal Title
Breast Cancer Research and Treatment
Journal Volume
179
Journal Issue
3
Journal Page Range
p. 671-676
ISSN
0167-6806
CODEN
BCTRD6

Optional Information

Copyright
Copyright (c) 2019 © The Author(s) 2019