Published February 1986 | Version v1
Journal article

The electronics for the Donner 600-Crystal Positron Tomograph

  • 1. Donner Lab. and Lawrence Berkeley Lab., Univ. of California, Berkeley, CA 94720

Description

The data acquisition system, designed for the Donner 600-Crystal Positron Tomograph, is described. Coincidence timing resolution of less than five nanoseconds full width at half maximum and data rates in excess of one million events per second are achieved by using high-speed emitter coupled logic circuits, first-in first-out memory to derandomize data flow, and parallel architecture to increase throughput. These data rates allow the acquisition of adequate transmission data in a reasonable amount of time. Good timing resolution minimizes accidental coincidences and permits data rates greater than 100,000 image-forming events per second for high-speed dynamic emission tomography. Additional scatter and accidental rejection are accomplished for transmission data by using an orbiting source and a look-up table for valid events. Calibration of this complex electronic system is performed automatically under computer control

Additional details

Publishing Information

Journal Title
IEEE Trans. Nucl. Sci.
Journal Volume
NS-33
Journal Issue
1
Series
IEEE Trans. Nucl. Sci.
Journal Page Range
vp.
ISSN
0018-9499
CODEN
IETNA

Conference

Title
IEEE nuclear science symposium.
Dates
23-25 Oct 1985.
Place
San Francisco, CA (USA).

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
17068509
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Resource subtype / Literary indicator
Conference
Descriptors DEI
CALIBRATION; DATA ACQUISITION SYSTEMS; IMAGE PROCESSING; LOGIC CIRCUITS; MEMORY DEVICES; ON-LINE CONTROL SYSTEMS; POSITRON COMPUTED TOMOGRAPHY; TIME RESOLUTION
Descriptors DEC
COMPUTERIZED TOMOGRAPHY; CONTROL SYSTEMS; ELECTRONIC CIRCUITS; EMISSION COMPUTED TOMOGRAPHY; ON-LINE SYSTEMS; RESOLUTION; TIMING PROPERTIES; TOMOGRAPHY

Optional Information

Secondary number(s)
CONF-851009--.