Published September 1992
| Version v1
Journal article
Study on detection efficiency of slow neutron proportional counter
Creators
- 1. Academia Sinica, Beijing, BJ (China). Inst. of Atomic Energy
Description
The detection efficiencies of typical slow neutron proportional counters and the correction coefficients for Maxwell spectra are deduced. For 3He and BF3 proportional counters, the measured detection efficiency of thermal neutron is consistent with the theoretic value satisfactorily. In the energy range above 0.1 eV, the theoretic values deduced from 3He (n, p) T reaction are less than that of the Monte Carlo Calculation, in which (n, n) reaction has been taken into account, and the experimental values. However, the experimental values are better consistent with the Monte Carlo Calculation
Additional details
Publishing Information
- Journal Title
- Nuclear Electronics and Detection Technology
- Journal Volume
- 12
- Journal Issue
- suppl.
- Series
- Proceedings of 6th national coference on nuclear electronics and nuclear detection technology (A).
- Journal Page Range
- p. 246-251.
- ISSN
- 0258-0934
- CODEN
- HDYUEC
Conference
- Title
- 6. national conference on nuclear electronics and nuclear detection technology.
- Dates
- 21-26 Sep 1992.
- Place
- Weihai (China).
INIS
- Country of Publication
- China
- Country of Input or Organization
- China
- INIS RN
- 26011576
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- BF3 COUNTERS; CORRECTIONS; EFFICIENCY; GEOMETRY; HE-3 COUNTERS; MONTE CARLO METHOD; NEUTRON SPECTRA; SLOW NEUTRONS
- Descriptors DEC
- BARYONS; CALCULATION METHODS; ELEMENTARY PARTICLES; FERMIONS; HADRONS; MATHEMATICS; MEASURING INSTRUMENTS; NEUTRON DETECTORS; NEUTRONS; NUCLEONS; PROPORTIONAL COUNTERS; RADIATION DETECTORS; SPECTRA