Published 1987
| Version v1
Report
Restricted
Development and exploitation of the slit method for the characterization of x-ray screen-film combinations
- 1. Physikalisch-Technische Bundesanstalt, Braunschweig (Germany, F.R.)
Description
For the determination of the modulation transfer function (MTF) of screen-film combinations which are used in medical x-ray diagnostics a measuring method has been developed: the screen-film combination is exposed to x-rays behind a thin slit and the modulation transfer function is calculated from the resulting rather broad slit images on the film. After solving many technical and photographic problems, the slit method provides high precision in the MTF determination. The only objection against this method is the necessary high dose variation which has to be provided by the x-ray machine
Files
Additional details
Publishing Information
- Imprint Pagination
- 50 p.
- Report number
- EUR--10964
INIS
- Country of Publication
- European Commission (EC), Brussels (Belgium)
- Country of Input or Organization
- European Commission (EC), Brussels (Belgium)
- INIS RN
- 19036952
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- DENSITOMETERS; PHOTOGRAPHIC FILMS; PHOTOGRAPHY; SCREENS; TRANSFER FUNCTIONS; X-RAY EQUIPMENT
- Descriptors DEC
- EQUIPMENT; FUNCTIONS; MEASURING INSTRUMENTS; PHOTOMETERS
Optional Information
- Contract/Grant/Project number
- Contract 1417/1/0/073/83/11-BCR-D(30)