Published 1986
| Version v1
Report
Open
The IPNS resonance detector spectrometer
Description
In the resonance detector method of neutron spectroscopy, a foil is placed in the scattered neutron beam and scattered neutrons having the resonance energy are detected by detecting the capture gammas resulting from the resonance absorption of the neutrons. A prototype resonance detector spectrometer called the Electron Volt Spectrometer (EVS) has been built and operated. The instrument is described, the current understanding of the background of the instrument is discussed, software developed to simulate the detector efficiency is described and compared with experimental results, and a test of the use of foil-thickness difference techniques to improve resolution is presented
Availability note (English)
MF available from INIS under the Report Number; Available from NTIS, PC A03/MF A01 as DE87004982.Files
18066125.pdf
Files
(589.0 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:1691243650b21770c58ec19337e76934
|
589.0 kB | Preview Download |
Additional details
Publishing Information
- Imprint Pagination
- 35 p.
- Report number
- CONF-8609234--3
Conference
- Title
- International collaboration on advanced neutron sources.
- Dates
- 22-26 Sep 1986.
- Place
- Villigen (Switzerland).
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 18066125
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- BACKGROUND RADIATION; COMPUTERIZED SIMULATION; EFFICIENCY; ENERGY RESOLUTION; FOILS; GAMMA CASCADES; GAMMA DETECTION; NEUTRON SPECTROMETERS; RESONANCE ABSORPTION; SPECIFICATIONS
- Descriptors DEC
- ABSORPTION; DETECTION; ENERGY-LEVEL TRANSITIONS; MEASURING INSTRUMENTS; NUCLEAR CASCADES; RADIATION DETECTION; RADIATIONS; RESOLUTION; SIMULATION; SPECTROMETERS