Published 1987 | Version v1
Book

Field trial of an automated quality assurance program for CT scanners

Description

A field trial of an automated quality assurance program for CT scanners utilizing a personal computer has been under way for several months. This paper reports on setting up the program and the initial results of automatically analyzing CT image quality parameters including spatial resolution, noise, uniformity, low contrast, CT linearity, section widths, and relates aspects of CT performance. Trends in short-term and long-term performance are analyzed against a backdrop of preventive maintenance records and overall system performance. A brief discussion is given of developments at both the voluntary and regulatory levels for a future CT quality assurance program

Additional details

Publishing Information

Publisher
Radiological Society of North America Inc.
Imprint Place
Oak Brook, IL (USA)
Imprint Title
Radiological Society of North America 73rd scientific assembly and annual meeting (Abstracts)
Journal Page Range
vp.

Conference

Title
73. scientific assembly and annual meeting of the Radiological Society of North America.
Dates
29 Nov - 4 Dec 1987.
Place
Chicago, IL (USA).

INIS