Published 2006 | Version v1
Miscellaneous Open

Validation of an activity optimization method for nuclear medicine in planar studies

Description

A method for optimizing the administered activity in Static Nuclear Medicine Studies is validated by comparison with ROC curve. Linear Discriminant analysis of image quality in gamma cameras was the applied statistical technique. The constructed linear discriminant function owns as dependent parameters, the differentiated levels of image quality obtained by observer's criterion. The independent parameters in the function were physical variables, as Signal-to Background ratios and Signal-to-Noise ratios. They were obtained from the selection of Regions of Interest in images obtained from a Jaszczak phantom, corresponding to lesion and background sites. The percentage of cases correctly classified by discriminant analysis was analyzed to grade the proposed discriminant method. The minimum value of the administered activity, which permits good image quality, (it means good results for the parameters selected by the discriminant function), can be proposed as an optimized value of activity for planar studies of Nuclear Medicine. The method was tested using images from a Jaszczak phantom, acquired under four activities (1088 MBq, 962 MBq, 740 MBq and 562 MBq) with a gamma camera equipped with a high resolution - low energy- parallel-hole collimator. The gamma camera was tested by a NEMA protocol. Image quality was graded by three expert observers who also developed a rated procedure which consist in analyzing the images for ROC analysis. Two of the six measured Background-to-Signal ratios were the parameters able to construct the linear discriminant function with high correlation respect to the observer criterion, from all the measured physical variables. The value of 740 MBq was the optimum after discriminant method application in this particular experiment. The results were coincident with the application of ROC-analysis. The optimal activity value obtained with the proposed discriminant procedure coincided with the activity value for which the area under the ROC curves was higher, the sensitivity was the best and the false-positives and false-negatives rates were the smallest. (Author)

Availability note (English)

Available from INIS in electronic form

Files

38030454.pdf

Files (59.3 kB)

Name Size Download all
md5:1beeddc5683179851a009d0f6bd0166e
59.3 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
7 p.
Report number
INIS-MX--2180

Conference

Title
1. american congress of the IRPA; 7. regional congress of radiological and nuclear safety; 24. annual meeting of the SMSR; 17. annual congress of the SNM
Original Conference Title
Retos y tendencias para el continente americano respecto a la proteccion radiologica, la seguridad nuclear y el ambiente
Dates
3-8 Sep 2006
Place
Acapulco, Gro. (Mexico)

INIS

Country of Publication
Mexico
Country of Input or Organization
Mexico
INIS RN
38030454
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Resource subtype / Literary indicator
Conference
Descriptors DEI
ACTIVITY LEVELS; COMPARATIVE EVALUATIONS; GAMMA CAMERAS; NUCLEAR MEDICINE; OPTIMIZATION; PHANTOMS; RADIOLOGY; SIGNAL-TO-NOISE RATIO; STATISTICS; VALIDATION
Descriptors DEC
CAMERAS; DIMENSIONLESS NUMBERS; EVALUATION; MATHEMATICS; MEDICINE; MOCKUP; NUCLEAR MEDICINE; STRUCTURAL MODELS; TESTING