Validation of an activity optimization method for nuclear medicine in planar studies
Creators
- Perez D, M.1
- Diaz R, O.2
- Farias L, F.3
- International Radiation Protection Association (IRPA) (France)
- Sociedad Mexicana de Seguridad Radiologica A.C. (SMSR), Mexico D.F. (Mexico)
- Sociedad Nuclear Mexicana (SNM), Mexico D.F. (Mexico)
- Organismo Internacional de Energia Atomica (OIEA), Vienna (Australia)
- Federacion de Radioproteccion de America Latina y el Caribe (FRALC), Zacatecas (Mexico)
- Organizacion Panamericana de la Salud (OPS), Washington, D.C. (United States)
- 1. Central University of Las Villas, CEETI, Camajuani Road Km 5.5, Santa Clara 54830 Villa Clara (Cuba)
- 2. Institute for Sciences and Advanced Technologies (Cuba)
- 3. Federal University of Pernambuco (Brazil)
Description
A method for optimizing the administered activity in Static Nuclear Medicine Studies is validated by comparison with ROC curve. Linear Discriminant analysis of image quality in gamma cameras was the applied statistical technique. The constructed linear discriminant function owns as dependent parameters, the differentiated levels of image quality obtained by observer's criterion. The independent parameters in the function were physical variables, as Signal-to Background ratios and Signal-to-Noise ratios. They were obtained from the selection of Regions of Interest in images obtained from a Jaszczak phantom, corresponding to lesion and background sites. The percentage of cases correctly classified by discriminant analysis was analyzed to grade the proposed discriminant method. The minimum value of the administered activity, which permits good image quality, (it means good results for the parameters selected by the discriminant function), can be proposed as an optimized value of activity for planar studies of Nuclear Medicine. The method was tested using images from a Jaszczak phantom, acquired under four activities (1088 MBq, 962 MBq, 740 MBq and 562 MBq) with a gamma camera equipped with a high resolution - low energy- parallel-hole collimator. The gamma camera was tested by a NEMA protocol. Image quality was graded by three expert observers who also developed a rated procedure which consist in analyzing the images for ROC analysis. Two of the six measured Background-to-Signal ratios were the parameters able to construct the linear discriminant function with high correlation respect to the observer criterion, from all the measured physical variables. The value of 740 MBq was the optimum after discriminant method application in this particular experiment. The results were coincident with the application of ROC-analysis. The optimal activity value obtained with the proposed discriminant procedure coincided with the activity value for which the area under the ROC curves was higher, the sensitivity was the best and the false-positives and false-negatives rates were the smallest. (Author)
Availability note (English)
Available from INIS in electronic formFiles
38030454.pdf
Files
(59.3 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:1beeddc5683179851a009d0f6bd0166e
|
59.3 kB | Preview Download |
Additional details
Publishing Information
- Imprint Pagination
- 7 p.
- Report number
- INIS-MX--2180
Conference
- Title
- 1. american congress of the IRPA; 7. regional congress of radiological and nuclear safety; 24. annual meeting of the SMSR; 17. annual congress of the SNM
- Original Conference Title
- Retos y tendencias para el continente americano respecto a la proteccion radiologica, la seguridad nuclear y el ambiente
- Dates
- 3-8 Sep 2006
- Place
- Acapulco, Gro. (Mexico)
INIS
- Country of Publication
- Mexico
- Country of Input or Organization
- Mexico
- INIS RN
- 38030454
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ACTIVITY LEVELS; COMPARATIVE EVALUATIONS; GAMMA CAMERAS; NUCLEAR MEDICINE; OPTIMIZATION; PHANTOMS; RADIOLOGY; SIGNAL-TO-NOISE RATIO; STATISTICS; VALIDATION
- Descriptors DEC
- CAMERAS; DIMENSIONLESS NUMBERS; EVALUATION; MATHEMATICS; MEDICINE; MOCKUP; NUCLEAR MEDICINE; STRUCTURAL MODELS; TESTING