Published September 2001
| Version v1
Journal article
Comparing the performance of mono- and biplane fluoroscopy systems in diagnostic and interventional neuroangiography using the dose-area product
- 1. National Hospital for Neurology and Neurosurgery, London (United Kingdom). Lysholm Dept. of Radiology
- 2. Dept. of Neuroradiology, Atkinson Morley's Hospital, London (United Kingdom)
- 3. Radiation Protection Centre, St George's Hospital, London (United Kingdom)
Description
Biplane digital imaging systems are favoured in neurointerventional practice since they are believed to contribute to safer, quicker and more efficacious procedures. Our aim was to establish if such equipment exposes patients to the same level of ionising radiation as monoplane systems. We reviewed monoplane and biplane fluoroscopy screening times and dose-area products (DAP) for 267 diagnostic cerebral angiograms and 56 neurointerventional cases. Significantly lower DAP for the latter were recorded on the biplane equipment, demonstrating an important reduction in patient exposure to radiation. There were no significant differences between the two systems for diagnostic cerebral angiography. (orig.)
Additional details
Publishing Information
- Journal Title
- Neuroradiology
- Journal Volume
- 43
- Journal Issue
- 9
- Journal Page Range
- p. 728-734
- ISSN
- 0028-3940
- CODEN
- NRDYAB
INIS
- Country of Publication
- Germany
- Country of Input or Organization
- Germany
- INIS RN
- 33025229
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BLOOD VESSELS; BRAIN; COMPARATIVE EVALUATIONS; DIAGNOSIS; FLUOROSCOPY; RADIATION DOSES; REVIEWS
- Descriptors DEC
- BIOMEDICAL RADIOGRAPHY; BODY; CARDIOVASCULAR SYSTEM; CENTRAL NERVOUS SYSTEM; DIAGNOSTIC TECHNIQUES; DOCUMENT TYPES; DOSES; EVALUATION; MEDICINE; NERVOUS SYSTEM; NUCLEAR MEDICINE; ORGANS; RADIOLOGY
Optional Information
- Notes
- With 3 figs., 3 tabs.