Published August 2018 | Version v1
Journal article

The relationship between applied energy and ablation zone volume in patients with hepatocellular carcinoma and colorectal liver metastasis

  • 1. University of Groningen, University Medical Center Groningen, Center for Medical Imaging North East Netherlands, Groningen (Netherlands)
  • 2. University of Groningen, University Medical Center Groningen, Department of Radiology, Groningen (Netherlands)
  • 3. University of Groningen, University Medical Center Groningen, Department of HPB Surgery, Groningen (Netherlands)

Description

To study the ratio of ablation zone volume to applied energy in computed tomography (CT)-guided radiofrequency ablation (RFA) and microwave ablation (MWA) in patients with hepatocellular carcinoma (HCC) in a cirrhotic liver and in patients with colorectal liver metastasis (CRLM). In total, 90 liver tumors, 45 HCCs in a cirrhotic liver and 45 CRLMs were treated with RFA or with one of two MWA devices (MWAA and MWAB), resulting in 15 procedures for each tumor type, per device. Device settings were recorded and the applied energy was calculated. Ablation volumes were segmented on the contrast-enhanced CT scans obtained 1 week after the procedure. The ratio of ablation zone volume in milliliters to applied energy in kilojoules was determined for each procedure and compared between HCC (RHCC) and CRLM (RCRLM), stratified according to ablation device. With RFA, RHCC and RCRLM were 0.22 mL/kJ (0.14-0.45 mL/kJ) and 0.15 mL/kJ (0.14-0.22 mL/kJ; p = 0.110), respectively. With MWAA, RHCC was 0.81 (0.61-1.07 mL/kJ) and RCRLM was 0.43 (0.35-0.61 mL/kJ; p = 0.001). With MWAB, RHCC was 0.67 (0.41-0.85 mL/kJ) and RCRLM was 0.43 (0.35-0.61 mL/kJ; p = 0.040). With RFA, there was no significant difference in energy deposition ratio between tumor types. With both MWA devices, the ratios were higher for HCCs. Tailoring microwave ablation device protocols to tumor type might prevent incomplete ablations. (orig.)

Availability note (English)

Available from: http://dx.doi.org/10.1007/s00330-017-5266-1

Additional details

Identifiers

Publishing Information

Journal Title
European Radiology
Journal Volume
28
Journal Issue
8
Journal Page Range
p. 3228-3236
ISSN
0938-7994
CODEN
EURAE3