Published January 2017 | Version v1
Journal article

Key comparison BIPM.RI(I)-K1 of the air-kerma standards of the SCK-CEN, Belgium and the BIPM in 60Co gamma radiation

  • 1. Bureau International des Poids et Mesures, F-92312 Sevres Cedex (France)
  • 2. Belgian Nuclear Research Centre - Studiecentrum voor Kernenergie - Centre d'Etude de l'Energie Nucleaire - SCK-CEN, Boeretang 200, 2400 Mol, (Belgium)

Description

A first key comparison of the standards for air kerma of the Laboratory for Nuclear Calibrations (LNK) from the Studiecentrum voor Kernenergie - Centre d'Etude de l'Energie Nucleaire (SCK-CEN), Belgium and of the Bureau International des Poids et Mesures (BIPM) was carried out in the 60Co radiation beam of the BIPM in September 2016. The comparison result, evaluated as a ratio of the SCK-CEN and the BIPM standards for air kerma, is 1.0021 with a combined standard uncertainty of 2.6 x 10-3. The results for an indirect comparison made at the same time are consistent with the direct results at the level of 7 parts in 104. The results are analysed and presented in terms of degrees of equivalence, suitable for entry in the BIPM key comparison database. (authors)

Additional details

Publishing Information

Journal Title
Metrologia
Journal Volume
54
Journal Issue
no.1A
Journal Page Range
p. 06004.1-06004.11
ISSN
0026-1394

Optional Information

Notes
12 refs.