Published August 2015 | Version v1
Journal article

A front-end readout Detector Board for the OpenPET electronics system

  • 1. Lawrence Berkeley National Laboratory, Berkeley, CA 94720 (United States)
  • 2. Fermi National Accelerator Laboratory, Batavia, IL 60510 (United States)

Description

We present a 16-channel front-end readout board for the OpenPET electronics system. A major task in developing a nuclear medical imaging system, such as a positron emission computed tomograph (PET) or a single-photon emission computed tomograph (SPECT), is the electronics system. While there are a wide variety of detector and camera design concepts, the relatively simple nature of the acquired data allows for a common set of electronics requirements that can be met by a flexible, scalable, and high-performance OpenPET electronics system. The analog signals from the different types of detectors used in medical imaging share similar characteristics, which allows for a common analog signal processing. The OpenPET electronics processes the analog signals with Detector Boards. Here we report on the development of a 16-channel Detector Board. Each signal is digitized by a continuously sampled analog-to-digital converter (ADC), which is processed by a field programmable gate array (FPGA) to extract pulse height information. A leading edge discriminator creates a timing edge that is ''time stamped'' by a time-to-digital converter (TDC) implemented inside the FPGA . This digital information from each channel is sent to an FPGA that services 16 analog channels, and then information from multiple channels is processed by this FPGA to perform logic for crystal lookup, DOI calculation, calibration, etc

Availability note (English)

Available from http://dx.doi.org/10.1088/1748-0221/10/08/T08002

Additional details

Publishing Information

Journal Title
Journal of Instrumentation
Journal Volume
10
Journal Issue
08
Journal Page Range
p. T08002
ISSN
1748-0221