Neutron detection channel with a counter based on the SNM-14 corona counter
Description
The block diagram is presented and performance characteristics of the neutron detection channel with a counter based on the SNM-14 corona counter are described. The channel sensitivity to neutrons of Pu-Be source is about 0.6 pulsesxcm2/neutron. A new way of testing the stability of performance of all the measurement line is proposed. It is based on using corona noise as a control pulse source. These devices and testing are developed for a dosymetric control system of the JINR high energy proton accelerator (''F'' installation)
Availability note (English)
MF available from INIS under the Report Number.Files
14772839.pdf
Files
(237.8 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:e98e90eda4387e24c5be552a1bb6f257
|
237.8 kB | Preview Download |
System files
(22.9 kB)
| Name | Size | Download all |
|---|
Additional details
Additional titles
- Original title (Russian)
- Канал детектирования нейтронов с датчиком на основе коронного счетчика SNM-14
Publishing Information
- Imprint Pagination
- 5 p.
- Report number
- JINR--13-82-743
INIS
- Country of Publication
- USSR
- Country of Input or Organization
- USSR
- INIS RN
- 14772839
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CORONA COUNTERS; COUNTING RATES; FLOWSHEETS; JINR CYCLOTRONS; NEUTRON DETECTION; NEUTRON SOURCES; NOISE; PERFORMANCE; PULSES; RADIATION MONITORING; SENSITIVITY; STABILITY
- Descriptors DEC
- ACCELERATORS; CYCLIC ACCELERATORS; CYCLOTRONS; DIAGRAMS; INFORMATION; ISOCHRONOUS CYCLOTRONS; MEASURING INSTRUMENTS; MONITORING; PARTICLE SOURCES; RADIATION DETECTION; RADIATION DETECTORS; RADIATION SOURCES
Optional Information
- Notes
- 4 refs.; 1 fig.; 1 tab.