Published November 2014 | Version v1
Miscellaneous Open

Verification of Co-60 blood irradiator mechanical engineering design

  • 1. Czech Technical University, Faculty of Nuclear Sciences and Physical Engineering, Prague (Czech Republic)

Description

Aim of this work is to verify the mechanical engineering design of the blood irradiator (from view of radiation protection) and estimation of the field homogeneity in the irradiation chamber. To estimate the radiation field around the blood irradiator a modeling approach have been chosen and Monte Carlo code MCNPX has been employed. Fluence-to-dose conversion factor from ICRP-74 have been used for the annual dose equivalent evaluation. With this results, the shielding analysis has been performed in the context of occupation factor at the service staff work place and the current legislation requirements. In homogeneity of the radiation field has been estimated on the basis of isodoses in the water phantom (imitate the irradiator object, e.g. blood unit) placed in the irradiation chamber.The results have been further discussed with regards to the additional optimization (cost, availability of a sufficient number of resources, etc.). (authors)

Files

46091455.pdf

Files (38.4 kB)

Name Size Download all
md5:0537309c321cf65004b3b42f8612970e
38.4 kB Preview Download
Part of:
XXXVI. Days of Radiation Protection. Book of Abstracts

Additional details

Publishing Information

Publisher
Environment, a.s.
Imprint Place
Nitra (Slovakia)
ISBN
978-80-89384-08-4
Imprint Title
XXXVI. Days of Radiation Protection. Book of Abstracts
Imprint Pagination
144 p.
Journal Page Range
p. 51
Report number
INIS-SK--2015-002

Conference

Title
36. Days of Radiation Protection Conference
Dates
10-14 Nov 2014
Place
Poprad (Slovakia)

Optional Information

Contract/Grant/Project number
Grants TACR ALFA-TA02010372; GACR-SGS12/200/OHK4/3T/14; RVO-68407700
Notes
Lector: Nikodemova, Denisa