Description

Some arguments are made regarding the medical usefulness of high spatial resolution in positron imaging, even if limited to small imaged volumes. Then the intrinsic limitations to spatial resolution in positron imaging are discussed. The project to build a small-volume, high resolution animal research prototype (SHARP) positron imaging system is described. The components of the system, particularly the detectors, are presented and brief mention is made of data acquisition and image reconstruction methods. Finally, some preliminary imaging results are presented; a pair of isolated point sources and 18F in the bones of a rabbit. Although the detector system is not fully completed, these first results indicate that the goals of high sensitivity and high resolution (4 mm) have been realized. (Auth.)

Additional details

Publishing Information

Publisher
North-Holland.
Imprint Place
Amsterdam (Netherlands)
ISBN
0 444 86534 9
Imprint Title
Positron annihilation
Imprint Pagination
1018 p.
Journal Page Range
p. 955-964.

Conference

Title
6. International conference on positron annihilation.
Dates
3 - 7 Apr 1982.
Place