Published May 1995 | Version v1
Journal article

Characterization and calibration of a large area beta scintillation detector for determination of Sr-90

  • 1. Westinghouse Savannah River Co., Aiken, SC (United States)

Description

A large area beta scintillation detector has been developed which is currently capable of determining Sr-90/Y-90 contamination in surficial soils. The detector system employs scintillating fiber optic arrays, with active dimensions approximately 15 cm wide by 100 cm long, both ends of which are coupled to multiple photomultiplier tubes (PMTs). Electronic processing includes coincidence requirements to optimize sensitivity and selectivity for the 2.28 MeV (maximum) beta particle from Y-90. Low energy beta particles and gamma rays are discriminated against using double ended and multi-layer coincidence requirements. The detector systems is personal-computer-software controlled and data restored in a format compatible with standard database software for ease of final data reduction. Experimental calibration studies have shown a linear response for Sr-90/Y-90 soil concentrations from 12 to over 500 pCi/g and a discrimination factor of 50 to 1 versus Cs-137. (author) 3 refs.; 4 figs

Additional details

Publishing Information

Journal Title
Journal of Radioanalytical and Nuclear Chemistry
Journal Volume
193
Journal Issue
1
Journal Page Range
p. 99-106.
ISSN
0236-5731
CODEN
JRNCDM

Conference

Title
3. International Conference on Methods and Applications of Radioanalytical Chemistry.
Acronym
MARC-III
Dates
10-16 Apr 1994.
Place
Kailua-Kona, HI (United States).