Published January 2011 | Version v1
Journal article

On the calibration of astigmatism particle tracking velocimetry for microflows

  • 1. Institute of Fluid Mechanics and Aerodynamics, Bundeswehr University Munich, 85577 Neubiberg (Germany)

Description

Astigmatism particle tracking velocimetry (APTV) is a method to determine three components (3C) of the velocity field in a volume (3D) using a single camera. The depth position of the particles is coded by optical distortions caused by a cylindrical lens in the optical setup. This technique is particularly suited for microfluidic applications as measurement errors due to spatial averaging and depth of correlation, typically encountered with μPIV approaches, are eliminated so that the measurement precision is enhanced. Unfortunately, the current state of the technique is limited by the small measurement region achievable with the current calibration procedures as well as by higher order image aberrations (Cierpka et al 2010 Meas. Sci. Technol. 21 045401). In order to extend the size of the measurement volume and to account for all image aberrations, a new intrinsic calibration procedure, based on the imaging function of the particles, is proposed in the paper at hand. It provides an extended measurement depth, taking into account all image aberrations. In this work, the calibration procedure was applied to a μPIV arrangement but could also be implemented on macroscopic experimental setups. The calibration procedure is qualified with synthetic data as well as Poiseuille flow in a straight rectangular micro-channel with a cross-sectional area of 200 ×  500 µm2. The three-dimensional velocity distribution of the whole channel was resolved via APTV with uncertainties of 0.9% and 3.7% of the centerline velocity, uc, for the in-plane and out-of-plane components, respectively. Further investigations using different cylindrical-lens focal lengths, magnifications and particle sizes provide information about achievable measurement depths and help to design and adapt the optimal system for the desired experiment

Availability note (English)

Available from http://dx.doi.org/10.1088/0957-0233/22/1/015401

Additional details

Identifiers

DOI
10.1088/0957-0233/22/1/015401;
PII
S0957-0233(11)64401-3;

Publishing Information

Journal Title
Measurement Science and Technology
Journal Volume
22
Journal Issue
1
Journal Page Range
[10 p.]
ISSN
0957-0233
CODEN
MSTCEP

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
45010651
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
ACCURACY; CALIBRATION; CAMERAS; DEPTH; ERRORS; IMAGES; LAMINAR FLOW; LENSES; MEASURING METHODS; PARTICLE SIZE; VELOCIMETERS; VELOCITY
Descriptors DEC
DIMENSIONS; FLUID FLOW; MEASURING INSTRUMENTS; SIZE