Published May 1979 | Version v1
Journal article

Monitoring ventricular function at rest and during exercise with a nonimaging nuclear detector

  • 1. Johns Hopkins Medical Institutions, Baltimore, MD

Description

A portable nonimaging device, the nuclear stethoscope, for measuring beat to beat ventricular time-activity curves in normal people and patients with heart disease, both at rest and during exercise, is being developed and evaluated. The latest device has several operting modes that facilitate left ventricular and background localization, measurement of transit times and automatic calculation and display of left ventricular ejection fraction. The correlation coefficient of left ventricular ejection fraction obtained with the device and with a camera-computer system was 0.92 in 35 subjects. During bicycle exercise the ejection fraction in 15 normal persons increased from 44 to 64% (P less than 0.001), whereas among 12 patients with heart disease it was unchanged in 5 and decreased in 7

Additional details

Publishing Information

Journal Title
Am. J. Cardiol.
Journal Volume
43
Journal Issue
5
Series
Am. J. Cardiol.
Journal Page Range
975-979
ISSN
0002-9149

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
12596108
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE; S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
CAMERAS; CARDIOVASCULAR DISEASES; COMPUTERS; DYNAMIC FUNCTION STUDIES; EXERCISE; HEART; MEASURING INSTRUMENTS; NUCLEAR MEDICINE; SPECIFICATIONS; TESTING
Descriptors DEC
BODY; CARDIOVASCULAR SYSTEM; DISEASES; MEDICINE; ORGANS