Monitoring ventricular function at rest and during exercise with a nonimaging nuclear detector
Creators
- 1. Johns Hopkins Medical Institutions, Baltimore, MD
Description
A portable nonimaging device, the nuclear stethoscope, for measuring beat to beat ventricular time-activity curves in normal people and patients with heart disease, both at rest and during exercise, is being developed and evaluated. The latest device has several operting modes that facilitate left ventricular and background localization, measurement of transit times and automatic calculation and display of left ventricular ejection fraction. The correlation coefficient of left ventricular ejection fraction obtained with the device and with a camera-computer system was 0.92 in 35 subjects. During bicycle exercise the ejection fraction in 15 normal persons increased from 44 to 64% (P less than 0.001), whereas among 12 patients with heart disease it was unchanged in 5 and decreased in 7
Additional details
Publishing Information
- Journal Title
- Am. J. Cardiol.
- Journal Volume
- 43
- Journal Issue
- 5
- Series
- Am. J. Cardiol.
- Journal Page Range
- 975-979
- ISSN
- 0002-9149
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 12596108
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- CAMERAS; CARDIOVASCULAR DISEASES; COMPUTERS; DYNAMIC FUNCTION STUDIES; EXERCISE; HEART; MEASURING INSTRUMENTS; NUCLEAR MEDICINE; SPECIFICATIONS; TESTING
- Descriptors DEC
- BODY; CARDIOVASCULAR SYSTEM; DISEASES; MEDICINE; ORGANS