Report on EUROMET.RI(I)-K1 and EUROMET.RI(I)-K4 (EUROMET project no. 813): Comparison of air kerma and absorbed dose to water measurements of 60Co radiation beams for radiotherapy
Creators
- Csete, I.1
- Leiton, A.G.2
- Sochor, V.3
- Lapenas, A.4
- Grindborg, J.E.5
- Jokelainen, I.6
- Bjerke, H.7
- Dobrovodsky, J.8
- Megzifene, A.9
- Hourdakis, C.J.10
- Ivanov, R.11
- Vekic, B.12
- Kokocinski, J.13
- Cardoso, J.14
- Buermann, L.15
- Tiefenboeck, W.16
- Stucki, G.17
- Van Dijk, E.18
- Toni, M.P.19
- Minniti, R.20
- McCaffrey, J.P.21
- Silva, C.N.M.22
- Kharitonov, I.23
- Webb, D.24
- Saravi, M.25
- Delaunay, F.26
- 1. National Office of Measures (OMH) - pilot laboratory and corresponding author (Hungary)
- 2. Research Centre for Energy, Environment and Technology (CMRI-CIEMAT) (Spain)
- 3. Czech Metrology Institute (CMI) (Czech Republic)
- 4. Latvian National Metrology Center (LNMC-RMTC) (Latvia)
- 5. Swedish Radiation Protection Authority (SSI) (Sweden)
- 6. Radiation and Nuclear Safety Authority (STUK) (Finland)
- 7. Norwegian Radiation Protection Authority (NRPA) (Norway)
- 8. Slovak Institute of Metrology (SMU) (Slovakia)
- 9. International Atomic Energy Agency, IAEA, Vienna (Austria)
- 10. Hellenic Atomic Energy Committee (HAEC-HIRCL) (Greece)
- 11. National Centre of Metrology (NCM) (Bulgaria)
- 12. Rudjer Boskovic Institute (IRB) (Croatia)
- 13. Central Office of Measures (GUM) (Poland)
- 14. Institute for Nuclear Technology (ITN-LMRIR) (Portugal)
- 15. Physikalisch Technische Bundesanstalt (PTB) (Germany)
- 16. Bundesamt fur Eich und Vermesungswesen (BEV) (Austria)
- 17. 17 Bundesamt fur Metrologie (METAS) (Switzerland)
- 18. NMi Van Swinden Laboratorium (NMi) (Netherlands)
- 19. ENEA-CR Istituto Nazionale di Metrologia delle Radiazioni Ionizzanti (ENEA) (Italy)
- 20. 20 National Institute of Standards and Technology (NIST) (United States)
- 21. National Research Council Canada (NRC) (Canada)
- 22. National Metrology Laboratory of Ionizing Radiation (LNMRI-IRD) (Brazil)
- 23. D I Mendeleyev Institute for Metrology (VNIIM) (RU)
- 24. Australian Radiation Protection and Nuclear Safety Agency (ARPANSA) (Australia)
- 25. National Atomic Energy Commission (CNEA-CAE) (Argentina)
- 26. Laboratoire National Henri Becquerel (LNE-LNHB) (France)
Description
The results of an unprecedented international effort involving 26 countries are reported. The EUROMET.RI(I)-K1 and EUROMET.RI(I)-K4 key comparisons were conducted with the goal of supporting the relevant calibration and measurement capabilities (CMC) planned for publication by the participant laboratories. The measured quantities were the air kerma (Kair) and the absorbed dose to water (Dw) in 60Co radiotherapy beams. The comparison was conducted by the pilot laboratory MKEH (Hungary), in a star-shaped arrangement from January 2005 to December 2008. The calibration coefficients of four transfer ionization chambers were measured using two electrometers. The largest deviation between any two calibration coefficients for the four chambers in terms of air kerma and absorbed dose to water was 2.7% and 3.3% respectively. An analysis of the participant uncertainty budgets enabled the calculation of degrees of equivalence (DoE), in terms of the deviations of the results and their associated uncertainties. As a result of this EUROMET project 813 comparison, the BIPM key comparison database (KCDB) will include eleven new Kair and fourteen new Dw DoE values of European secondary standard dosimetry laboratories (SSDLs), and the KCDB will be updated with the new DoE values of the other participant laboratories. The pair-wise degrees of equivalence of participants were also calculated. In addition to assessing calibration techniques and uncertainty calculations of the participants, these comparisons enabled the experimental determinations of NDw/NKair ratios in the 60Co gamma radiation beam for the four radiotherapy transfer chambers. (authors)
Availability note (English)
Available from doi: <http://dx.doi.org/10.1088/0026-1394/47/1A/06012Additional details
Identifiers
Publishing Information
- Journal Title
- Metrologia
- Journal Volume
- 47
- Journal Issue
- no.3sup.06012
- Journal Page Range
- p. 1-100
- ISSN
- 0026-1394
- CODEN
- MTRGAU
INIS
- Country of Publication
- France
- Country of Input or Organization
- France
- INIS RN
- 42010340
- Subject category
- S07: ISOTOPES AND RADIATION SOURCES;
- Descriptors DEI
- AIR; CALIBRATION STANDARDS; COBALT 60; DATA COVARIANCES; DOSE RATES; INTERLABORATORY COMPARISONS; IONIZATION CHAMBERS; KERMA; RADIOTHERAPY; WATER
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; COBALT ISOTOPES; FLUIDS; GASES; HYDROGEN COMPOUNDS; INTERMEDIATE MASS NUCLEI; INTERNAL CONVERSION RADIOISOTOPES; ISOMERIC TRANSITION ISOTOPES; ISOTOPES; MEASURING INSTRUMENTS; MEDICINE; MINUTES LIVING RADIOISOTOPES; NUCLEAR MEDICINE; NUCLEI; ODD-ODD NUCLEI; OXYGEN COMPOUNDS; RADIATION DETECTORS; RADIOISOTOPES; RADIOLOGY; STANDARDS; THERAPY; YEARS LIVING RADIOISOTOPES
Optional Information
- Notes
- 17 refs.