Measuring Activity Concentrations Using a High Purity Germanium Detector: Extended Source Efficiency Correction and Optimization of Sample Position

  • 1. Grupo de Física Nuclear, Universidad Nacional de Colombia, Bogotá (Colombia)

Description

An experiment using several calibrated gamma-ray sources located at different positions around a high purity germanium detector was conducted. The efficiency calibration curve for each position was obtained. More than 15 gamma-ray energies for each position point were used. The efficiency calibration curves facilitate the determination of radionuclide concentrations in an extended source such as NORM. These curves are useful for the measurement of a sample that is small enough to be considered as a point source, in order to obtain the position which produces the larger number of counts for the gamma-ray energy of interest. Preliminary results of the resulting efficiency curves and their interpretation are presented. (author)

Part of:
Naturally Occurring Radioactive Material (NORM VIII). Proceedings of an International Symposium

Additional details

Publishing Information

Publisher
IAEA
Imprint Place
Vienna (International Atomic Energy Agency (IAEA))
ISBN
978-92-0-107618-2
Imprint Title
Naturally Occurring Radioactive Material (NORM VIII). Proceedings of an International Symposium
Imprint Pagination
424 p.
Series
Proceedings Series
Journal Page Range
p. 351-355
ISSN
0074-1884

Conference

Title
8. International Symposium on Naturally Occurring Radioactive Material
Acronym
NORM VIII
Dates
18-21 Oct 2016
Place
Rio de Janeiro (Brazil)

Optional Information

Notes
Poster presentation; 4 refs., 5 figs.
Secondary number(s)
STI/PUB--1832