Published 2002 | Version v1
Journal article

Are exhaust fans effective in reducing contamination to staff from the Technegas ventilation process?

  • 1. Liverpool Hospital, NSW (Australia). Department of Nuclear Medicine and Clinical Ultrasound

Description

Full text: To investigate the effectiveness of a Nederman exhaust fan (Edward Keller Industrial products Australia) in removing aerosol contamination released from the Technegas generator. A yellow impermeable gown was worn by the technologist administering the Technegas as per departmental protocol. The exhaust fan was used during administration on alternate day. After each ventilation scan the gown was placed into a clear plastic bag and sealed with tape. A new gown was used for each ventilation scan. The gowns were surveyed for contamination using a Geiger counter, then imaged for 5 minutes using a Picker 2000XP dual head camera. This process was repeated for each ventilation scan Factors such as patient compliance, breathing difficulties, age and sex were documented for each study. Twenty patients were studied 11 - without the exhaust fan and 9 with. The use of the exhaust fan has the potential to significantly reduce contamination from Technegas. Routine use should be considered to maintain the As Low As Reasonably Achievable (ALARA) principle. Copyright (2002) The Australian and New Zealand Society of Nuclear Medicine Inc

Availability note (English)

Available in abstract form only, full text entered in this record

Additional details

Publishing Information

Journal Title
ANZ Nuclear Medicine
Journal Volume
33
Journal Issue
suppl
Journal Page Range
p. 6
ISSN
1324-1435

Conference

Title
32. Annual Scientific Meeting of the Australian and New Zealand Society of Nuclear Medicine
Dates
4-8 May 2002
Place
Cairns, QLD (Australia)

Optional Information

Notes
1 tabs.