Investigation of buildup dose from electron contamination of clinical photon beams
- 1. Department of Physics and Cyclotron Laboratory, Harvard University, Cambridge, Massachusetts 02138
Description
The contribution made by contaminating electrons present in a clinical photon beam to the buildup dose in a polystyrene phantom has been calculated and compared to measurements. A Monte Carlo technique was employed. The calculation was divided into two parts. First, the accelerator treatment head was simulated in detail using the EGS--PEGS electromagnetic shower code. Then, information obtained from these calculations was used to compute dose curves in a polystyrene phantom. Two cases were considered, one in which both electrons and photons were incident upon the phantom, and another in which electrons were eliminated from the incident beam. Results of these calculations agree with recent experimental findings. A decrease in buildup dose as well as a shift in d/sub max/ was observed when electrons were eliminated from the beam
Additional details
Publishing Information
- Journal Title
- Med. Phys.
- Journal Volume
- 10
- Journal Issue
- 1
- Series
- Med. Phys.
- Journal Page Range
- 18-24
- ISSN
- 0094-2405
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 14768533
- Subject category
- S07: ISOTOPES AND RADIATION SOURCES;
- Descriptors DEI
- COMPUTERIZED SIMULATION; CONTAMINATION; DEPTH DOSE DISTRIBUTIONS; DOSE-RESPONSE RELATIONSHIPS; DOSIMETRY; E CODES; ELECTRONS; GEOMETRY; LINEAR ACCELERATORS; MEV RANGE; MONTE CARLO METHOD; PHANTOMS; PHOTON BEAMS; POLYSTYRENE; SHOWERS
- Descriptors DEC
- ACCELERATORS; BEAMS; COMPUTER CODES; DISTRIBUTION; ELEMENTARY PARTICLES; ENERGY RANGE; FERMIONS; LEPTONS; MATHEMATICS; MOCKUP; ORGANIC COMPOUNDS; ORGANIC POLYMERS; POLYMERS; POLYOLEFINS; POLYVINYLS; RADIATION DOSE DISTRIBUTIONS; SIMULATION; SPATIAL DISTRIBUTION; SPATIAL DOSE DISTRIBUTIONS; STRUCTURAL MODELS