Measurement and analysis of CT number linearity

  • 1. Ningbo Center for Diseases Control and Prevention, Ningbo (China)

Description

Objective: To explore the evaluation method for CT number linearity by measuring CT numbers. Methods: The model CTP401 of Catphan500 phanton was adopted to measure CT numbers of four objective substances, then CT number Linearity, contrast scales, and uncertainties were calculated. Results: Statistically significant differences in CT numbers were found between the two reconstruction algorithms (P<0.001), The contrast scale of sharp reconstruction algorithm was 1.90 x 10-4 cm-1 · Hu-1, and smooth reconstruction algorithm was 1.79 x 10-4 cm-1 · Hu-1. Conclusions: Qualified CT number linearity is important to image quality control in CT scanners, Reconstruction algorithms and uncertainties must be marked in test reports of CT number linearity. (authors)

Additional details

Publishing Information

Journal Title
Chinese Journal of Radiological Health
Journal Volume
18
Journal Issue
3
Journal Page Range
p. 289-291