Published 1979 | Version v1
Miscellaneous

Design and fabrication of a self-powered neutron detector

Description

Self powered neutron detectors are becoming more and more popular in reactor instrumentation. A fast response detector of this type was made at the Reactor Division, La Reina Nuclear Center in Santiago. Cobalt wire was the emitter, teflon the insulator and a stainless steel tubing was the collector. The overall dimensions of the detector are 6 mms diameter and 700 mms length. The irradiation tests, carried out at the Center's 5 Mw research reactor showed a very reasonably linear relation between current supplied by the detector and thermal neutron flux, over a range extending from 1010 to 1013 n/cm2 x seg. These tests also showed a good agreement between calculated and measured current. The models used for the calculation of current are fully explained and they include some improvements over those that have been published recently. An important conclusion for the case of the cobalt detectors is that the wire's diameter must be at least 1 mm. in order to have a neutron induced current bigger than the parasitic components generated by indirect processes. Calculations for other emitters such as vanadium, silver and rhodum are also included. (EC)

Availability note (English)

Available from Comision Chilena de Energia Nuclear, Santiago (Chile).

Additional details

Additional titles

Original title (Spanish)
Diseno y construccion de un detector de neutrones autoalimentado

Publishing Information

Imprint Pagination
149 p.

INIS

Country of Publication
Chile
Country of Input or Organization
Chile
INIS RN
11555476
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Thesis, Non-conventional Literature
Descriptors DEI
FABRICATION; NEUTRON DETECTORS; NUMERICAL SOLUTION; SELF-POWERED NEUTRON DETECTORS
Descriptors DEC
MEASURING INSTRUMENTS; RADIATION DETECTORS; SELF-POWERED DETECTORS