Published December 1999 | Version v1
Journal article

Moving-bed infusion-tracking three-dimensional MR angiography by applying conventional CP body array coil

  • 1. Osaka City Univ. (Japan). Hospital

Description

We devised a new method of Mobi-track MRA by applying the conventional CP body array coil and our own sliding table. This method was applied for MR angiography of lower leg arteries combined with a subtraction technique. The performance of our method, which uses the CP body array coil, was examined in regard to misregistration of difference images, signal-to-noise ratio, contrast-to-noise ratio, and signal detectability, for comparison with that of the method using the body coil. Experimental results showed that our new method was superior to the method using the body coil. In conclusion, we consider our new method, which employs the CP body array coil and sliding table, to be very useful for three-dimensional MR angiography of lower leg arteries. (author)

Additional details

Publishing Information

Journal Title
Nippon Hoshasen Gijutsu Gakkai Zasshi
Journal Volume
55
Journal Issue
12
Journal Page Range
p. 1226-1231
ISSN
0369-4305