Gray matter-white matter contrast on spin-echo T1-weighted images at 3 T and 1.5 T: a quantitative comparison study
Creators
- 1. Kyoto University Graduate School of Medicine, Department of Diagnostic Imaging and Nuclear Medicine, Kyoto (Japan)
- 2. Kyoto University Graduate School of Medicine, Human Brain Research Center, Kyoto (Japan)
- 3. National Institute of Neuroscience, National Center of Neurology and Psychiatry, Department of Cortical Function Disorders, Kyoto (Japan)
Description
Discrepancies exist in the literature regarding contrast between gray and white matter on spin-echo (SE) T1-weighted MR imaging at 3 T. The present study quantitatively assessed differences in gray matter-white matter contrast on both single- and multi-slice SE T1-weighted imaging between 3 and 1.5 T. SE T1-weighted sequences with the same parameters at both 3 and 1.5 T were used. Contrast-to-noise ratio (CNR) between gray and white matter (CNRGM-WM) was evaluated for both frontal lobes. To assess the effects of interslice gap, multi-slice images were obtained with both 0 and 25% interslice gap. Single-slice CNRGM-WM was higher at 3 T (17.66 ± 2.68) than at 1.5 T (13.09 ± 2.35; P < 0.001). No significant difference in CNRGM-WM of multi-slice images with 0% gap was noted between 3 and 1.5 T (3T, 8.61 ± 2.55; 1.5T, 7.43 ± 1.20; P > 0.05). Multi-slice CNRGM-WM with 25% gap was higher at 3T (12.47 ± 3.31) than at 1.5 T (9.73 ± 1.37; P < 0.001). CNRGM-WM reduction rate of multi-slice images with 0% gap compared with single-slice images was higher at 3T (0.47 ± 0.13) than at 1.5 T (0.38 ± 0.09; P = 0.02). CNRGM-WM on single-slice SE T1-weighted imaging and CNRGM-WM on multi-slice images with 25% interslice gap were better at 3 T than at 1.5 T. The influence of multi-slice imaging on CNRGM-WM was significantly larger at 3T than at 1.5 T. (orig.)
Availability note (English)
Available from: http://dx.doi.org/10.1007/s00330-007-0688-9Additional details
Identifiers
Publishing Information
- Journal Title
- European Radiology
- Journal Volume
- 17
- Journal Issue
- 11
- Journal Page Range
- p. 2921-2925
- ISSN
- 0938-7994
- CODEN
- EURAE3
INIS
- Country of Publication
- Germany
- Country of Input or Organization
- Germany
- INIS RN
- 38115909
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BRAIN; NMR IMAGING; SPIN ECHO
- Descriptors DEC
- BODY; CENTRAL NERVOUS SYSTEM; DIAGNOSTIC TECHNIQUES; NERVOUS SYSTEM; ORGANS