A timing discriminator for channel plate electron multipliers
Creators
- 1. Paris-11 Univ., 91 - Orsay (France). Inst. de Physique Nucleaire
Description
A constant fraction discriminator used with channel plate electron multipliers (CPEM) provides extremely good time resolution, yet it has a simple design and does not need delicate adjustment. The results obtained depend on the type of CPEM apparatus which is used. For electron detection (β conversion electrons) a resolution of 300 ps can be achieved for a system using two discriminators and two pairs of channel plate multipliers. For a similar double system this resolution becomes 65 ps for the detection of secondary electrons produced by the scattering of 252Cf fission fragments in a carbon foil. The same type of discriminator can be used in systems employing fast photomultiplier tubes. (Auth.)
Additional details
Publishing Information
- Journal Title
- Nucl. Instrum. Methods
- Journal Volume
- 161
- Journal Issue
- 2
- Series
- Nucl. Instrum. Methods.
- Journal Page Range
- 331-337
- ISSN
- 0029-554X
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- Netherlands
- INIS RN
- 10485116
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- BETA DETECTION; CALIFORNIUM 252; ELECTRON CHANNELING; ELECTRON DETECTION; ELECTRON MULTIPLIER DETECTORS; FISSION FRAGMENTS; MICROCHANNEL ELECTRON MULTIPLI; PHOTOMULTIPLIERS; PULSE CIRCUITS; PULSE DISCRIMINATORS; SPECIFICATIONS; TIME RESOLUTION
- Descriptors DEC
- ACTINIDE NUCLEI; ALPHA DECAY RADIOISOTOPES; CALIFORNIUM ISOTOPES; CHANNELING; CHARGED PARTICLE DETECTION; DISCRIMINATORS; ELECTRON MULTIPLIERS; ELECTRON TUBES; ELECTRONIC CIRCUITS; EVEN-EVEN NUCLEI; HEAVY NUCLEI; ISOTOPES; MEASURING INSTRUMENTS; NUCLEAR FRAGMENTS; NUCLEI; PHOTOTUBES; RADIATION DETECTION; RADIATION DETECTORS; RADIOISOTOPES; RESOLUTION; TIMING PROPERTIES; YEARS LIVING RADIOISOTOPES