Neutron transport in a clinical linear accelerator bunker: comparison of materials for reducing the photo-neutron dose at the maze entrance
Creators
Description
Megavoltage photons above 10 MeV used in external beam radiotherapy lead to a significant photo-neutron fluence, which must be taken into account in bunker design (IPEM, 1997, Report 75, The design of radiotherapy treatment room facilities). This work describes Monte-Carlo simulations of such neutrons for a proposed bunker, which is to house a 15 MV accelerator. Neutron fluence spectra and absorbed dose due to neutrons and neutron-capture photons were scored at the accelerator iso centre and at the maze entrance for mono-energetic neutron sources of 0.5, 3 and 6 MeV. The reduction in neutron and photon dose at the maze entrance, achieved by cladding concrete maze walls with either wood, polyethylene or a commercially available plastic, was determined.
Additional details
Identifiers
- PII
- S0969806X01003541;
Publishing Information
- Journal Title
- Radiation Physics and Chemistry (1993)
- Journal Volume
- 61
- Journal Issue
- 3-6
- Journal Page Range
- p. 633-634
- ISSN
- 0969-806X
- CODEN
- RPCHDM
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- Iran, Islamic Republic of
- INIS RN
- 43042207
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE; S38: RADIATION CHEMISTRY, RADIOCHEMISTRY AND NUCLEAR CHEMISTRY;
- Descriptors DEI
- LEAD; MONTE CARLO METHOD; PHOTONEUTRONS; PHOTONS; RADIOTHERAPY; SIMULATION
- Descriptors DEC
- BARYONS; BOSONS; CALCULATION METHODS; ELEMENTARY PARTICLES; ELEMENTS; FERMIONS; HADRONS; MASSLESS PARTICLES; MEDICINE; METALS; NEUTRONS; NUCLEAR MEDICINE; NUCLEONS; PHOTONUCLEONS; RADIOLOGY; THERAPY
Optional Information
- Copyright
- Copyright (c) 2001 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.
- Notes
- This record replaces 34052412