Published 1996 | Version v1
Book

A count rate model for PET and its application to an LSO HR PLUS scanner

  • 1. TRIUMF, Vancouver, British Columbia (Canada)

Description

We present a count rate model for PET. Considering a standard 20 x 20 cm phantom in the field-of-view of a cylindrical septaless tomograph, the model computes the acceptance to prompt and random events from simple geometric considerations. Dead time factors at all stages of a typical event acquisition architecture are calculated from specified processing clock cycles. Validations of the model's predictions against the measured performances of the ECAT-953B and the EXACT HR PLUS are presented. The model is then used to investigate the benefit of using detectors made of LSO in the EXACT HR PLUS scanner geometry. The results indicate that in replacing BGO by the faster LSO, one can count on an increase of the peak noise-equivalent-count rate by a factor 2.2. This gain will be achieved by using a 5 nsec coincidence window, buckets operating on a 128 nsec clock cycle, and a front-end data acquisition that can sustain a total rate of 2.9 MHz

Additional details

Publishing Information

Publisher
IEEE Service Center.
Imprint Place
Piscataway, NJ (United States)
Imprint Title
1996 IEEE nuclear science symposium - conference record. Volumes 1, 2 and 3
Imprint Pagination
2138 p.
Journal Page Range
p. 1186-1190.

Conference

Title
Institute of Electrical and Electronic Engineers (IEEE) nuclear science symposium and medical imaging conference.
Dates
2-9 Nov 1996.
Place
Anaheim, CA (United States).

Optional Information

Secondary number(s)
CONF-961123--.