Cryostat for annealing spectra measurements with short rise time to the annealing temperature

  • 1. Democritus Nuclear Research Center, Athens (Greece)

Description

Abstract A simple cryostat for annealing spectra measurements in the temperature range 78 to 700 °K is described and its specifications are given. The rise time at the annealing temperature is chosen equal to 1.5 minutes, giving thus, a practically square shape of the temperature pulse. An analogue valve regulates automatically a stream of cold nitrogen gas and maintains the temperature of the samples constant within ±0.05 °C. The rise time varies from 1 min, or even less, to about 7 min, when there is no control of temperature through the cold gas.

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Effects
Journal Volume
10
Journal Issue
4
Series
Radiat. Eff.
Journal Page Range
215-218
ISSN
0033-7579

INIS

Country of Publication
United Kingdom
Country of Input or Organization
United Kingdom
INIS RN
3016808
Subject category
S30: DIRECT ENERGY CONVERSION;
Resource subtype / Literary indicator
No Abstract
Descriptors DEI
ACTIVATION ENERGY; ANALOG SYSTEMS; ANNEALING; CONTROL SYSTEMS; COOLING; CRYOSTATS; FURNACES; GAS FLOW; IRRADIATION; LABORATORY EQUIPMENT; LATTICES; NITROGEN; POINT DEFECTS; PULSES; RADIATION EFFECTS; RELIABILITY; SOLIDS; SPECTRA; TEMPERATURE; TIME DEPENDENCE; VALVES
Descriptors DEC
CRYOGENICS; DEFECTS; EXCITATION; PERFORMANCE

Optional Information

Notes
Updated automatically by Metadata and Full-Text Enrichment Agent