Published September 11, 1998
| Version v1
Journal article
An analyzer for pulse-interval times to study high-order effects in the processing of nuclear detector signals
Creators
Description
An electronic device to measure interval time density distributions of subsequent pulses in nuclear detectors and their electronics is described. The device has a pair-pulse resolution of 10 ns and 25 ns for 3 subsequent input signals. The conversion range is 4096 channels and the lowest channel width is 10 ns. Counter dead times, single and in series were studied and compared with the statistical model. True count rates were obtained from an exponential fit through the interval-time distribution
Additional details
Identifiers
- PII
- S0969804397100276;
Publishing Information
- Journal Title
- Applied Radiation and Isotopes
- Journal Volume
- 49
- Journal Issue
- 9-11
- Journal Page Range
- p. 1099-1105
- ISSN
- 0969-8043
- CODEN
- ARISEF
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 34013542
- Subject category
- S07: ISOTOPES AND RADIATION SOURCES;
- Descriptors DEI
- COUNTING RATES; DEAD TIME; PULSE RISE TIME; STATISTICAL MODELS; TIME INTERVAL ANALYZERS; TIME RESOLUTION
- Descriptors DEC
- MATHEMATICAL MODELS; MEASURING INSTRUMENTS; RESOLUTION; TIMING PROPERTIES
Optional Information
- Copyright