An analyzer for pulse-interval times to study high-order effects in the processing of nuclear detector signals

Description

An electronic device to measure interval time density distributions of subsequent pulses in nuclear detectors and their electronics is described. The device has a pair-pulse resolution of 10 ns and 25 ns for 3 subsequent input signals. The conversion range is 4096 channels and the lowest channel width is 10 ns. Counter dead times, single and in series were studied and compared with the statistical model. True count rates were obtained from an exponential fit through the interval-time distribution

Additional details

Identifiers

PII
S0969804397100276;

Publishing Information

Journal Title
Applied Radiation and Isotopes
Journal Volume
49
Journal Issue
9-11
Journal Page Range
p. 1099-1105
ISSN
0969-8043
CODEN
ARISEF

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
34013542
Subject category
S07: ISOTOPES AND RADIATION SOURCES;
Descriptors DEI
COUNTING RATES; DEAD TIME; PULSE RISE TIME; STATISTICAL MODELS; TIME INTERVAL ANALYZERS; TIME RESOLUTION