Published 1981 | Version v1
Report

Ion implanted Si detector for radiometric investigations of liquids

Description

Silicon detectors for α-radiometric investigations of liqiuds has been tested in the 239Pu labelled solution. A tendency to sorption has been studied of different structural materials of silicon semiconductor diodes as well as polystyrene - the main part of plastic scintillators. It is proved that the sorption on the surface of detectors has been immesurable, while for polystyrene the residual activity of 2-10 decays per minxcm2 has been obtained. Contact with acid causes no change in the current-voltage characteristics, whereas α-spectra result in a slight amplitude change

Additional details

Additional titles

Original title (Russian)
Ионноимплантированный кремниевый детектор для измерения активности жидкостей

Publishing Information

Imprint Title
Ways of improving semiconductor detectors
Journal Page Range
p. 47-49.
Report number
KIYI--81-24

Optional Information

Notes
3 refs.; 2 figs.; 1 tab. Imprint:Puti sovershenstvovaniya poluprovodnikovykh detektorov.