Published March 1980 | Version v1
Journal article

The influence of charge collection characteristics on HPGe detector timing performance

  • 1. Queen's Univ., Kingston, Ontario (Canada). Dept. of Physics

Description

The expected response of a constant fraction timing discriminator to pulse shapes from a coaxial HPGe dtector has been calculated for several choices of charge drift length and impurity concentration over a range of calculated detector bias, and also for both possible electron and hole drift directions. Optimum timing performance is obtained with impurity concentrations of the order of 1010 cm-3 and when electrons drift towards the detector centre. (orig.)

Additional details

Publishing Information

Journal Title
Nucl. Instrum. Methods
Journal Volume
169
Journal Issue
3
Series
Nucl. Instrum. Methods.
Journal Page Range
465-468
ISSN
0029-554X

INIS